Comunicación

AN OPTOELECTRONIC SYSTEM FOR MEASURING THERANGE OF MOTION IN HEALTHY VOLUNTEERS: A CROSS-SECTIONAL STUDY

Autores:

FRANCISCO MEDINA MIRAPEIX1, RODRIGO MARTIN SAN AGUSTIN2, GERMAN CANOVAS AMBIT3, JOSÉ ANTONIO GARCÍA VIDAL1, MARIANO GACTO SANCHEZ4, MARIA PILAR ESCOLAR REINA1

Afiliaciones:

(1) FISIOTERAPIA Y DISCAPACIDAD, IMIB, España
(2) PROFESOR UNIVERSIDAD DE VALENCIA, 46010, España (Comunidad Valenciana)
(3) FISIOTERAPEUTA, 30100, España (Región de Murcia)
(4) PROFESOR UNIVERSIDAD DE GIRONA, 17190, España (Cataluña)

Publicación asociada:

PMID: 31443368
Referencia: Medina-Mirapeix F, Martín-San Agustín R, Cánovas-Ambit G, García-Vidal JA,Gacto-Sánchez M, Escolar-Reina P. An Optoelectronic System for Measuring theRange of Motion in Healthy Volunteers: A Cross-Sectional Study. Medicina(Kaunas). 2019 Aug 22;55(9). pii: E516. doi: 10.3390/medicina55090516. PubMedPMID: 31443368.
ISSN: 1010-660X
Revista: Medicina-Lithuania
Factor de impacto (2017): 2.6
Cuartil: 3

Comunicación:

Antecedentes:

Within the clinical evaluation of multiple pathologies of the lower limb, the measurement of range of motion (ROM) of its joints is fundamental. To this end, there are various tools, from the goniometer as a reference to more recent devices such as inclinometry-based applications, photo capture applications, or motion capture systems. This study aimed to assess the validity, intra-rater, and inter-rater reliability of the VeloFlex system (VS), which is a new camera-based tool designed for tracking joint trajectories and measuring joint ROM

Métodos:

Thirty-five healthy volunteers (16 females; aged 18–61 years) participated in this study. All participants were assessed on two separate occasions, one week apart. During the first assessment session, measurements were obtained using a goniometer and the VS, whereas, in the second session, only the VS was used. In each assessment session, nine active movements were examined. For each movement, three trials were tested, and the mean of these three measures was used for analysis. To evaluate the concurrent validity and agreement, the Pearson correlation coecient (r) and Bland-Altmann plots were used. Intra-rater and inter-rater reliability were evaluated using intra-class correlation (ICC), standard error of measurement (SEM), and minimal detectable change (MDC).

Resultados:

Both devices showed excellent correlations for all movements (r ranged from 0.992 to 0.999). The intra-rater reliability of the VS was excellent (ICC ranged from 0.93 to 0.99), SEMs ranged from 0.53% to 2.61% and the MDC ranged from 0.68 to 3.26. The inter-rater reliability of the VS was also excellent (ICC ranged from 0.88 to 0.98), SEMs ranged from 0.81% to 4.76% and the MDC ranged from 2.27 to 4.42

Conclusiones:

The Veloflex system is a valid and reliable tool for the measurement of ROM of lower limb joints in healthy subjects.


Dirección

Campus de Ciencias de la Salud
Carretera Buenavista s/n, 30120 El Palmar
Murcia, España

Ver en OpenStreetMap

Ver en Google Maps

Teléfonos

+34 868885229
+34 868885239
+34 868885249